In the medical device world, Usability Engineering (IEC 62366 – The application of Usability Engineering or its 60601 counterpart IEC 60601-1-6) is a required tool within the validation of a medical device. Regardless of whether it’s software, active, or IDV, Usability Engineering helps us understand how the device (or system) performs in real-world applications.

Risk Assessment and Usability

Usability Engineering draws from the risk principles outlined in ISO 14971 and identifies hazardous situations that users may encounter while using the device. However, it encompasses much more than just risk identification.

Understanding Key Concepts

Many organizations struggle with the concepts of “hazard-related use scenario” and “hazards.” Let’s clarify these with practical examples:

  • The risk assessment process follows standard medical device protocols
  • Special attention to EN ISO 14971 vs. ISO 14971 for EEA markets

Practical Example: Smoke Alarm System

Technical Controls

Consider a typical smoke alarm:

  • A system failure might be an insufficiently loud siren
  • ISO 12239 specifies compliant fire system development
  • Controls involve designing to these requirements and verification testing

Usability Engineering Focus

Usability examines the human-device interface, which can include:

  • Sounds (buzzers, beeps)
  • Visual indicators (lights, text, symbols)
  • Tactile feedback
  • Training materials

We test by placing users in controlled hazardous scenarios to evaluate their response.

Implementing Effective Controls

Example Hazardous Scenario

Situation: User doesn’t understand the alarm meaning and fails to act

Potential Controls:

  • Training materials
  • Reference documentation
  • Clear visual indicators

Note: We exclude “unforeseeable abnormal use” (malicious intent) from consideration, focusing only on normal and foreseeable abnormal use conditions.

Proper Verification Methods

Avoid simply checking if controls exist – this only confirms implementation, not effectiveness.

Effective Verification Process:

  1. Provide training and materials to users
  2. Simulate alarm activation
  3. Observe user responses

Measure success by evaluating how identifiable, understandable, and usable the controls were in preventing the hazardous situation.

Why Usability Engineering Matters

This simplified example demonstrates how usability engineering applies to medical devices. Effective usability studies provide critical feedback for device development and improvement.